Friday, May 27, 2022

FDA recall of floor lifts

ArjoHuntleigh Polska is recalling the Sara Plus floor lift following
several complaints of smoke and/or flames coming out of the lift. When
the battery is low, the device's printed circuit board may overheat,
leading to smoke or fire. If this situation occurs, anyone using the
lift or near it could be injured, including smoke inhalation and/or
burns.

The FDA has identified this as a Class I recall, the most serious type
of recall. Use of these devices may cause serious injuries, serious
health consequences, or death.